| | SLO | ENG | Cookies and privacy

Bigger font | Smaller font

Show document Help

Title:VALIDACIJA STERILIZACIJE - DOLOČITEV ROKA UPORABE STERILIZIRANEGA MATERIALA
Authors:ID Salajko, Damir (Author)
ID Goršek, Andreja (Mentor) More about this mentor... New window
ID Pečar, Darja (Comentor)
Files:.pdf VS_Salajko_Damir_2013.pdf (2,24 MB)
MD5: B82BE47C064B55E06D35BE55099D4DF9
PID: 20.500.12556/dkum/ce3ab988-4830-4dba-aaed-8f24f84fc4cf
 
Language:Slovenian
Work type:Bachelor thesis/paper
Typology:2.11 - Undergraduate Thesis
Organization:FKKT - Faculty of Chemistry and Chemical Engineering
Abstract:Namen diplomskega dela je bil z validacijo dokazati, da material (oprema), steriliziran po predpisanem postopku v parnem sterilizatorju - avtoklavu, ohrani sterilnost tudi po 30 dnevih hranjenja. V sklopu raziskav smo pripravili pakete z deli opreme, ki so nujno potrebni v proizvodnem procesu in jih vstavili v sterilizacijske posode poleg materiala (opreme), ki se uporablja v proizvodnem procesu. Te pakete smo nato izpostavili sterilizaciji s paro v avtoklavu. Po ohladitvi paketov na sobno tempeaturo smo izbrane dele steriliziranih paketov hranili v proizvodnih prostorih razreda čistosti D. Po 30 dnevih hranjenja smo na teh delih steriliziranih paketov določali sterilnost. Rezultati v diplomskem delu kažejo, da je ustrezno sterilizirane in zapakirane pakete materiala (opreme) možno hraniti 30 dni, ne da bi bila pri tem kakorkoli ogrožena njegova kvaliteta oz. sterilnost.
Keywords:validacija, sterilizacija, avtoklav, biološki indikatorji, distribucija temperature, rok uporabe, oprema
Place of publishing:Maribor
Publisher:[D. Salajko]
Year of publishing:2013
PID:20.500.12556/DKUM-40381 New window
UDC:614.48:615.014.4(043.2)
COBISS.SI-ID:17473302 New window
NUK URN:URN:SI:UM:DK:ESZ0AMWP
Publication date in DKUM:22.05.2013
Views:3106
Downloads:441
Metadata:XML DC-XML DC-RDF
Categories:KTFMB - FKKT
:
Copy citation
  
Average score:(0 votes)
Your score:Voting is allowed only for logged in users.
Share:Bookmark and Share



Hover the mouse pointer over a document title to show the abstract or click on the title to get all document metadata.

Secondary language

Language:English
Title:VALIDATION OF STERILIZATION - DETERMINATION OF STERILIZATION MATERIAL 'S USAGE TIME
Abstract:The purpose of our work was validation of sterilization which demonstrate that material (equipment) sterilized according to the prescribed procedure in a steam steriliser - autoclave maintains the sterility even after 30 days of storage. In the context of this thesis, we have prepared packages with items of equipment that are necessary in the production process and inserted it into sterilization containers loaded with the material (equipment) used in the production process. These packets were exposed to steam sterilization in an autoclave and than stored in the production area with cleanliness grade D. After 30 days of storage sterilized packages were analyzed to determine sterility. The results in this thesis show that properly sterilized and packed material (equipment) can be stored for 30 days, without being in any way compromised its quality or sterility.
Keywords:validation, sterilisation, autoclave, biological indicators, temperature distribution, shelf life, equipment