Title:Nizkocenovni sistem za merjenje oksigenacije kožnih, podkožnih in možganskih tkiv s pomočjo bližnje infrardeče spektroskopije
Authors:ID Zemljič, Sašo (Author)
ID Holobar, Aleš (Mentor) More about this mentor... New window
ID Cigale, Boris (Comentor)
Files:.pdf MAG_Zemljic_Saso_2017.pdf (3,62 MB)
MD5: C68ECCD5D0BD0A400097107C0217887B
 
Language:Slovenian
Work type:Master's thesis/paper
Typology:2.09 - Master's Thesis
Organization:FERI - Faculty of Electrical Engineering and Computer Science
Abstract:Bližnja infrardeča spektroskopija omogoča neinvazivno spremljanje oksigenacije tkiv. Zaradi optičnih lastnosti človeškega tkiva in vsebovanih kromofor, je svetloba bližnje infrardeče regije posebej primerna za spremljanje koncentracij oksi- in deoksihemoglobina. V magistrskem delu zasnujemo in izdelamo sistem za opravljanje meritev stopenj oksigenacije s pomočjo bližnje infrardeče spektroskopije. Izdelana naprava sestoji iz optoelektronskega senzorja in krmilnega vezja. Optoelektronski senzor sestavljajo dvobarvna svetleča dioda, ki oddaja svetlobo z valovnima dolžinama 750 in 850 nm ter par različno oddaljenih fotodiod. Za analogno digitalno pretvorbo signalov optoelektronskega senzorja skrbi krmilno vezje, ki zajete signale pošilja na osebni računalnik, kjer jih dekomponiramo in izračunamo relativne spremembe stopnje oksigenacije. Delovanje izdelanega sistema smo preverili z večkratnimi meritvami somatske in cerebralne oksigenacije zdravega preiskovanca.
Keywords:bližnja infrardeča spektroskopija, merjenje oksigenacije, somatska oksigenacija, cerebralna oksigenacija
Place of publishing:Maribor
Publisher:[S. Zemljič]
Year of publishing:2017
PID:20.500.12556/DKUM-65923 New window
UDC:[004:616-71]:615.835(043.2)
COBISS.SI-ID:20703510 New window
NUK URN:URN:SI:UM:DK:ZAH1C1MJ
Publication date in DKUM:03.07.2017
Views:1608
Downloads:236
Metadata:XML DC-XML DC-RDF
Categories:KTFMB - FERI
:
Copy citation
  
Average score:(0 votes)
Your score:Voting is allowed only for logged in users.
Share:Bookmark and Share



Hover the mouse pointer over a document title to show the abstract or click on the title to get all document metadata.

Secondary language

Language:English
Title:A low-cost system for nirs-based assessment of skin, adipose and brain tissue oxygenation
Abstract:Near-infrared spectroscopy supports non-invasive monitoring of tissue oxygenation. Due to the optical properties of human tissue and its contained chromophores, the light of near infrared region is especially suitable for monitoring concentrations of oxy- and deoxy hemoglobin. In this master's thesis we describe the design and implementation of a system for measuring oxygenation levels using near-infrared spectroscopy. This device comprises of an optoelectronic sensor and control circuit. Optoelectronic sensor is composed of a bi-color light emitting diode which emits light with wavelengths of 750 and 850 nm and a couple of differently spaced photodiodes. The analog to digital conversion of the optoelectronic sensor’s signal is carried out on the control circuit, which sends signals to the personal computer where they are decomposed and processed by specific algorithms to calculate the relative change in oxygenation level. We tested the designed system by performing multiple measurements of somatic and cerebral oxygenation on a healthy young subject.
Keywords:near-infrared spectroscopy, measurement of oxygenation, somatic oxygenation, cerebral oxygenation