Title:Simulacije priprave, delovanja in napak sistemov za dovajanje kisika z visokimi pretoki
Authors:ID Ivanović, Ilijana (Author)
ID Križmarić, Miljenko (Mentor) More about this mentor... New window
Files:.pdf MAG_Ivanovic_Ilijana_2023.pdf (1,93 MB)
MD5: 9AF1BE2D648CB3DC6864CB5F529A82CA
 
Language:Slovenian
Work type:Master's thesis/paper
Typology:2.09 - Master's Thesis
Organization:FZV - Faculty of Health Sciences
Abstract:Uvod: Za aplikacijo kisika lahko uporabljamo sisteme z različnimi pretoki, zato je potrebno ustrezno znanje o kisiku in sistemih, ki jih lahko uporabljamo. Nosna kanila z visokimi pretoki je dokaj nov inovativen sistem, ki se uporablja pri bolnikih, ki dihajo spontano in potrebujejo visoke koncentracije kisika. Metode: S pomočjo simulacije sistema za visoke pretoke smo simulirali pripravo, delovanje sistema in pogoste napake na sistemu. V eksperimentalnem delu smo uporabili sistem za dovajanje kisika z visokimi pretoki WILAmed. Za ogrevanje in vlaženje plinov smo uporabili enoto AIRcon Gen2 in Laerdal Airway Management Trainer za simulacijo dihalnega sistema. Rezultati: Dobljeni rezultati kažejo, da lahko simuliramo pripravo medicinskih plinov, sistema in napak na sistemu. Prav tako nam je uspelo preveriti ustrezne tlake medicinskih plinov v inštalacijskem sistemu in sprožiti alarm. Pri simulaciji napak smo se osredotočili na pogoste opozorilne alarme in slikovno prikazane napake na zaslonu ter jih uspešno simulirali. Razprava in zaključek: Višji pretoki bolje ustrezajo zahtevam pacientov, zmanjšujejo anatomski mrtvi prostor in zagotavljajo rahlo pozitiven tlak v zgornjih dihalnih poteh. Neželeni učinki teh sistemov so epistaksa, nelagodje v nosu in suhost. Potrebnih je več raziskav, ki bi natančno opredelile pozitivne in negativne lastnosti ter jih primerjale z neinvazivno ventilacijo.
Keywords:aplikacija kisika, visoki pretoki kisika, zdravstvena nega, neinvazivna ventilacija
Place of publishing:Maribor
Publisher:[I. Ivanović]
Year of publishing:2023
PID:20.500.12556/DKUM-86193 New window
UDC:004.94:615.816:616-083(043.2)
COBISS.SI-ID:176714499 New window
Publication date in DKUM:11.12.2023
Views:629
Downloads:71
Metadata:XML DC-XML DC-RDF
Categories:FZV
:
Copy citation
  
Average score:(0 votes)
Your score:Voting is allowed only for logged in users.
Share:Bookmark and Share



Hover the mouse pointer over a document title to show the abstract or click on the title to get all document metadata.

Licences

License:CC BY-NC-ND 4.0, Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International
Link:http://creativecommons.org/licenses/by-nc-nd/4.0/
Description:The most restrictive Creative Commons license. This only allows people to download and share the work for no commercial gain and for no other purposes.
Licensing start date:22.10.2023

Secondary language

Language:English
Title:Simulations of preparation, operation and failure of high-flow oxygen supply systems
Abstract:Introduction: For the application of oxygen, we can use systems with different flow rates, so it is necessary to have adequate knowledge about oxygen and the systems that can be used. The high-flow nasal cannula is a fairly new innovative system used in spontaneously breathing patients who require high concentrations of oxygen. Methods: With the help of a high-flow system simulation, we simulated the preparation and operation of the system, and common system errors. The WILAmed high-flow oxygen delivery system was used. We used the AIRcon Gen2 unit for heating and humidifying gases, and the Laerdal Airway Management Trainer for simulating the respiratory system. Results: The obtained results show that we can simulate the preparation of medical gases, of the system, and system errors. We also managed to check the appropriate medical gas pressures in the installation system and trigger the alarm. When simulating errors, we focused on the frequent warning alarms and the graphically displayed errors on the screen, and successfully simulated them. Discussion and Conclusion: Higher flows better meet patient requirements, reduce anatomic dead space, and provide slightly positive upper airway pressure. The side effects of these systems include epistaxis, nasal discomfort and dryness. More research is needed to precisely define the positive and negative features, and compare them with non-invasive ventilation.
Keywords:oxygen application, high oxygen flows, nursing, non-invasive ventilation